Recovery duration is the time
the sensor switches from mea-
suring calibration gas readings
to measuring process readings
without affecting the analog
output or triggering of alarms.
5-12 | Dycor CG1000-RTP Oxygen Analyzer

Recovery Duration

Define a recovery time.
Defining a recovery time is necessary so that the analyzer can return to
reading process gases after reading calibration gases without affecting the
analog outputs or the triggering of alarms.
This menu option applies to both calibrations and verifications.
NOTE
1. From the Calibrate key menu, select Set Clock.
2. Select Recovery Duration from the Set Clock menu option.
3. Enter the recovery time, then press ENTER. If you want to use the
currently displayed recovery time, press ENTER without entering any
text.
Cal Recovery = xx:yy
Enter new time: Mn:Sc