For verify operations, select one of the following:
Select one of these options, your choice will be confirmed on the display:

Set Current Mode Option

Define either a 0-to-20 or 4-to-20 mA analog output range for the selected
analog output port.
This output is dependent on both the range of readings you define (see
the Set Current Range section), and on the function you track on the
analog output.
Setting the current mode
1. After selecting an analog output port, from the Analog Range Key,
select Set Current Mode.
2. Select 4-to-20 mA Mode or 0-to-20 mA Mode.
Hold during verify
or
Track during verify
Out #1 will TRACK
during Verify
or
Out #1 will HOLD
during Verify
Out # y is set for
4 - 20 ma
or
Out # y is set for
0 - 20 ma
Controller / User Interface
|
4-11