4
As you put the furnace insulation cover back on the furnace, place the
thermocouple and furnace wires attached to the furnace through the
hole in the furnace insulation cover.
5
Reinstall the plate that fastens the furnace insulation cover to the fur-
nace by reversing the actions taken in Step 5 of removing the furnace.
6
Reconnect the thermocouple wires to TB2, TC+ and TC- on the back-
plane board. Yellow wire - TC+, Red wire -TC-.
7
Reconnect the furnace wires to TB1, AC and AC on the backplane
board. Polarity not required when reconnecting these wires.
8
Reconnect the cell assembly as described in the "Cell Replacement"
section.
9
Reattach the power plug to the analyzer.
10 Turn on the analyzer.
11 Recalibrate the analyzer.
|