A7.4.3 Configure And Trigger Test Gas Insert - Ametek OXYvisor Installation, Operation & Maintenance Manual

Optical oxygen analyzer for use with bosx optical sensors
Table of Contents

Advertisement

RELAY
REGISTERS
Relay Test Number
Relay Test Position
Relay Test Status
Table A.21 - Relay registers

A7.4.3 Configure and Trigger Test Gas Insert

Follow the steps below to configure and trigger the Test Gas Insert feature of the OXYvisor. If you have already
setup the Test Gas Insert previously, via the software or the HMI, you can proceed to steps 3 and 4 to activate
and deactivate this feature manually.
Before configuring the Test Gas Insert, ensure that the auto calibration has been enabled [Address- 40131 (1
to activate and 2 to deactivate)].
1.
To set the test gas insert time, set the value in minutes between 1-100 to address 40405
2.
Set the test gas insert output to either Zero Gas (0X01) or Span Gas (0X02) to address 40407
3.
Activate test gas insert (0x01), address 44009.
4.
Deactivate/stop the test gas insert before specified time, set value (0x02) to address 44009.
REGISTER NAME FOR
TEST GAS INSERT
Time Period
Define Output
Activate/Deactivate
Table A.22 - Configure and Trigger Test Gas Insert
A7.4.3 Configure and Trigger Test Gas Insert
The Auto-Cal feature in the OXYvisor has 4 different schedules. Each schedule can be activated/deactivated,
set up for 1st calibration time and calibration interval individually. First Schedule = [0], Second Schedule = [1],
Third Schedule = [2], Fourth Schedule = [3]
Follow the steps below to setup and trigger the Auto-Cal feature of the OXYvisor. If you have already setup
the Auto-Cal previously or via the software or the HMI, you can proceed to steps 15 and 16 to activate and
deactivate this feature manually. [Address- 44000 (write 4 to activate and 5 to deactivate)].
1.
Set the auto calibration type (Register Address: 40130). The value can be either Dry Gas (0X01) or
Humid Gas(0x02).
2.
Turn off auto calibration (Register Address: 40131). The value should be set to (0x02).
3.
Deactivate each calibration schedule (Register Address: 40132, 40142, 40152, 40162). The value
should be set to (0x02).
4.
Set the calibration interval for each schedule (Register Address: 40138, 40148, 40158, 40168). The
value should be between 0 – 5184000 which is 60days in seconds. For example, calibration interval of 1
calibration a day is 86400 seconds. Calibration interval of once a week is 604800 seconds. The value to write
to the register is in seconds.
5.
Set the first calibration time for each schedule(Register Address : 40140, 40150, 40160, 40170). The
value should be between the current time in seconds (UTC time) + the UTC time within the next 60 days.
For example, if the current date and time is February 13th, 2018 – 5:00PM, the UTC value in seconds is
78
Installation, Operation and Maintenance Manual
FIRST REGISTER
MINUS
OFFSET
44010
40000
44011
40000
44012
40000
FIRST
MINUS
REGISTER
OFFSET
40408
40000
40410
40000
44009
40000
OXYvisor Optical Oxygen Analyzer
FUNCTION CODE
DATA TYPE
6
Uint16
6
Uint16
6
Uint16
FUNCTION CODE
DATA TYPE
16
Float
6
Uint16
6
Uint16
VALUE TO WRITE
1,2,3 or 4 for
corresponding relay
0=Open; 1=Close
1=Activate;
2=Deactivate
VALUE TO WRITE
1-100 (in mins)
Zero Gas (0X01) or
Span Gas (0X02)
Start (0X01) or
Stop
(0X02)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents