4-12 | Dycor CG1000-RTP Oxygen Analyzer

Output Filtering Option

Filter out quick transitions in readings to allow a smoother analog output.
Setting output filtering
1. After selecting an analog output port, from the Analog Range Key
select Output Filtering.
2. Enter an output filtering value between 1 and 100 where XX is the cur-
rent filter value, then press ENTER.
Note that 100 equals no filtering, and 1 equals the highest possible
filter value. Note that you should enter 01 to enter 1; 07 to enter 7; 70
for 70, etc.
The default value is 100 (no filtering).
NOTE

Select Function Option

Select what type of information the analog output port should track. Choic-
es are:
Oxygen
Cell temperature
Thermocouple millivolts (T/C mV)
Cell millivolts (cell mV)
Selecting the function
1. After selecting an analog output port, from the Analog Range menu,
choose Select Function.
2. Select what you want the analog output port to track.
The CG1000-RTP prompts you with the Set Current Range menu option
where you can choose the range of readings that the 0 to 20 mA or 4 to 20
mA analog output represents (see the Set Current Range section in this
chapter for details).
100 = FAST
1 = SLOW
Fltr = XX New #?