Defining alarm set points
1. From the Alarm menu, select Alarm Set Points.
Once you select the Alarm Set Points menu option, the system alerts
you that Alarms 1 and 2 are reserved for the watchdog and service
alarms, and are not software selectable. Press Enter to continue to the
NOTE
configurable alarms.
2. If Alarm 3 is tracking oxygen, select whether you want to enter the
Alarm 3 set point using a percent or PPM range. If Alarm 3 is tracking
flow, set the high and low flow limits.
3. You will be prompted to define the Alarm 3 set point value:
4. Enter the Alarm 3 set point value, then press ENTER (just press EN-
TER if you don't want to change the current value).
5. Select whether you want to enter the Alarm 4 set point using a percent
or PPM range.
You are prompted to enter the Alarm 4 set point value:
6. Enter the Alarm 4 set point value.
7. Press Cancel or ENTER to exit this menu option.
Alarm #3 = xx
New Value?
Alarm #4 = xx
New Value?
Controller / User Interface
|
4-15