Configure Verify

Define the verify gas value and the acceptable range of readings for a
verify.
To set your verify gas value and acceptable verify range, follow these
steps:
1.
From the Calibrate key menu, select Configure Verify.
2.
Select whether you want to use a % or ppm gas for the verify opera-
tion.
Enter the verify gas value:
3.
Press Enter. If you want to keep the current verify gas value un-
changed, press Enter without entering a value.
4.
Enter the acceptable verify range. For example, if you enter a verify
gas range of 1%, and the verify gas value is 10%, then the acceptable
range is between 11% and 9%.
NOTE
Verify Gas = xx
New Value?
If during a verify operation (Initiate Verify menu option) the system
finds that the verify gas is not within the acceptable range, a Verify
Failure message will appear on the bottom line of the display, and the
Service Alarm will trip. This message will be cleared if a successful
verify is run (the verify gas is within the range specified), or a calibra-
tion is run.
|
Calibration
5-11