Ametek Dycor CG1000-RTP User Manual page 9

Table of Contents

Advertisement

SPECIAL WARNINGS AND INFORMATION
EQUIPMENT USED IN CLASS I, DIVISION 2 HAZARDOUS LOCATIONS
This Equipment is Suitable for Use in Class I, Division 2, Groups ABCD, T4 or Non-Hazardous
Areas Only.
All input and output wiring must be in accordance with Class I, Division 2 wiring methods (NEC
Sec 501.4(b) or CEC 18-152) and in accordance with the authority having jurisdiction.
If the CG1000-RTP is to be powered by a source of 24 VDC other than that supplied by AME-
TEK, the power source's output must be isolated from hazardous mains voltages using double
or reinforced insulation which has a minimum dielectric strength of 2300 VAC. When the
CG1000-RTP is used in a Class I, Division 2 area, this external power source must be located in a
general-purpose area or be Division 2-approved.
Division 2 stand-alone Analyzer is not supplied with the 24 VDC power sup-
ply option.)
Explosion Hazard - Substitution of Components May Impair Suitability for
Class I, Division 2.
Risque d'explosion - La substitution de composants peut rendre ce ma-
teriel inacceptable pour les emplacements de Classe I, Division 2.
Explosion Hazard - Do Not Disconnect Equipment Unless Power Has Been
Switched Off or the Area is Known to be Non-Hazardous.
Risque d'explosion - Avant de déconnecter l'équipement, coupez le cou-
rant où vous assurez que l'emplacement est designé non dangereux.
| ix