5-6 | Dycor CG1000-RTP Oxygen Analyzer
If your O
reading on the far left of the display has stabilized, yet is
2
not near the setpoint value on the right of the display, check the verify
gas value you entered and ensure that this value matches the cal gas
NOTE
cylinder for the verify gas you applied to the calibration inlet port on
the back of the analyzer.
If you specified a recovery time, the system will display the recovery time
remaining.
8. Set the 3-way valve that delivers the verify gas to OFF.
9. Set the 3-way valve on the back of the analyzer to GAS IN.
When returning to reading the sample gas, be sure to again adjust the
flow rate to 150 sccm (± 5 sccm).
NOTE
Recovery xx:xx
System Verifying