Table of Contents

Advertisement

Quick Links

Dycor CG1000-RTP
User Manual
Thermox
150 Freeport Road
PN 90355VE, Rev. N
Pittsburgh, PA 15238

Advertisement

Table of Contents
loading

Summary of Contents for Ametek Dycor CG1000-RTP

  • Page 1 Dycor CG1000-RTP User Manual Thermox 150 Freeport Road PN 90355VE, Rev. N Pittsburgh, PA 15238...
  • Page 2: Offices

    © 2012 AMETEK This manual is a guide for the use of the Dycor CG1000-RTP. Data herein has been verified and validated and is believed adequate for the intended use of this instrument. If the instrument or procedures are used for purposes over and above the capabilities specified herein, confirmation of their validity and suitability should be obtained;...
  • Page 3: Table Of Contents

    Contents OFFICES ......................... ii European Standards Information ............... x CHAPTER 1 CG1000-RTP Oxygen Analyzer Basic Elements of the Sensor ................1-1 The Oxygen Measuring Cell ................1-2 Hydrocarbons ....................1-3 Proper Sensor Operations ................1-4 Technical Support ....................1-5 CHAPTER 2 Specifications CHAPTER 3 Installation Mechanical Installation ..................3-2 PPM Measurement Suggestions ..............3-6 Electrical Installation ..................3-7...
  • Page 4 AC Power Checks..................6-7 Furnace Checks ....................6-7 Cell Checks ....................6-7 General Troubleshooting ...................6-8 CHAPTER 7 Parts Replacement Cell Replacement....................7-2 Furnace Replacement ..................7-5 Replacing the Transformer ................7-8 Replacing the Backplane Board ...............7-9 Replacing Display/Keypad Modules .............7-10 iv | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 5 APPENDIX A Serial Communications Communication Protocol ................. A-1 Command Message Format ................A-2 General Serial Communication Commands ..........A-3 Development Serial Communication Commands ........A-4 Command List ....................A-5 Responses ......................A-6 Variable Table ..................... A-7 APPENDIX B Menu Option Charts Setup Key Menu ....................
  • Page 6 Grounding Instrument grounding is mandatory. Performance specifications and safety protection are void if instrument is operated from an improperly grounded power source. Verify ground continuity of all equipment before applying power. vi | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 7 Achtung - Heiße Oberfläche Environmental Information (WEEE) This AMETEK product contains materials that can be reclaimed and recycled. In some cases the product may contain materials known to be hazardous to the environment or human health. In order to prevent the release of harmful sub- stances into the environment and to conserve our natural resources, AMETEK recommends that you arrange to recycle this product when it reaches its “end of...
  • Page 8 In general, for optimum protection against high frequency transients and other dis- turbances, do not allow installation of this Analyzer where its unshieled I/O and DC circuits are physically mixed with AC mains or any other circuit that could induce viii | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 9 SPECIAL WARNINGS AND INFORMATION EQUIPMENT USED IN CLASS I, DIVISION 2 HAZARDOUS LOCATIONS This Equipment is Suitable for Use in Class I, Division 2, Groups ABCD, T4 or Non-Hazardous Areas Only. Division 2 stand-alone Analyzer is not supplied with the 24 VDC power sup- ply option.) Explosion Hazard - Substitution of Components May Impair Suitability for Class I, Division 2.
  • Page 10: European Standards Information

    RF Radiated (pulsed) ENV 50141 RF Induced EN 61000-4-8 Magnetic Emissions using: EN 55011 Radiated, Class B and with the low voltage directive 73/23/EEC: EN 61010-1 (IEC 1010-1), Safety Requirement for Electrical Equip- ment x | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 11 Process photometric analyzers, process moisture analyzers, and sampling systems are warranted to perform the intended measurement, only in the event that the customer has supplied, and AMETEK has accepted, valid sample stream composition data, process conditions, and electrical area classification prior to order acknowledgment.
  • Page 12 This page intentionally left blank. xii | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 13: Chapter 1 Cg1000-Rtp Oxygen Analyzer

    CG1000-RTP OXYGEN ANALYZER Follow these safety precautions when working on the sensor: Disconnect AC power from the analyzer before performing any main- tenance or troubleshooting. The furnace and furnace insulation cover are hot during normal oper- ation (up to 500°F/260°C inside the furnace insulation cover). Allow analyzer components to cool for at least an hour before working inside the analyzer.
  • Page 14: The Oxygen Measuring Cell

    The oxygen content is then determined from the Nernst equation: where R and F are constants, T is absolute temperature, and O and O the oxygen partial pressures on either side of the cell. 1-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 15: Hydrocarbons

    Figure 1-1. Zirconium oxide cell principle of operation. For measuring oxygen in non-combustibles gases, the calibration of an analyzer is obtained from the formula: AT = 48.0 at 695°C 20.9% E = A * T * Log Unk% AT = 54.4 at 824°C Where A is a constant, T is the cell temperature on an absolute scale (°C + 273) and O Unk% is the unknown oxygen concentration of the gas to be...
  • Page 16: Proper Sensor Operations

    Allow at least one hour after closing the sensor door for readings to stabilize. Allow more time after a cold start-up. • If calibrating in the PPM range, calibration gas should be in cylinders made of aluminum, not carbon steel. 1-4 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 17: Technical Support

    Technical Support AMETEK/Thermox is committed to providing the best technical support in the industry. If you need service or application assistance, please call AMETEK at (412) 828-9040, or your local AMETEK/Thermox representative. Before you call the factory for technical support, run test gases and record...
  • Page 18 This page intentionally left blank. 1-6 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 19 SpecificationS Sensor operating Range 0.1 PPM O to 100% O Percent: ±2% of reading or .05% O absolute, whichever is greater. accuracy PPM: ± 2% of reading or .5 PPM O absolute, whichever is greater. Response time < 5 seconds at 150 sccm over one decade. Percent: ±...
  • Page 20 0.5A, 30V, 10VA max. noninductive load, AC or DC Watchdog timer and service alarms. System test for A/D, RAM, EEPROM and keypad. Display line 4 reserved for full-text error and Diagnostics diagnostic messages. Twenty-entry exception log for automatically detected system events. 2-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 21: Chapter 3 Installation

    INSTALLATION This chapter shows you how to install your CG1000-RTP oxygen analyzer. It includes the following: • Inspect Shipping Contents • Mechanical Installation Sample gas requirements Sample gas set-up Calibration gas set-up PPM measurement suggestions • Electrical Installation Current output connections Alarm contact connections RS-232 communications Select 115/230 Volt operation...
  • Page 22: Mechanical Installation

    O-ring seal. The sample gas inlet pressure into the CG1000-RTP must be less than 3 PSIG. If the sample contains particulates greater than 15 microns, install a filter on the sample gas inlet. Figure 3-1. Rear of CG1000-RTP analyzer. 3-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 23 Sample Gas Connection close Valve 1 and Valve 2. set the pressure of P1 on the pressure con- troller (on the vacuum generator) to zero (Figure 3-2). set the three-way valve on the back of the analyzer to OFF (Figure 3-1). connect the sample gas from Valve 1 to the sample gas 1/8”...
  • Page 24 10 (for example, if zero gas is 1.0%, span gas must be 10% or higher. If zero gas is .1%, span NOTE gas should be 1% or higher). 3-4 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 25 Figure 3-3 Calibration Gas Set-Up set the 3-way valve on the back of the analyzer to OFF. connect the calibration, span and zero gases to the 3-way valve. 1/8” using 1/8” tubing, connect the outlet of the 3-way valve to a 1/8” compres- tubing sion T-fitting.
  • Page 26: Ppm Measurement Suggestions

    The amount of oxygen reduction is dependent on the type of hydrocarbons present in the sample gas.  If measuring PPM gases, use a packless valve to switch between the sample gas and calibration gases. NOTE 3-6 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 27: Electrical Installation

    Electrical Installation current output connections  Alarm contact connections  Rs-232 communications  select 115/230 volt operation  Current Output Connections current output connections are labeled as follows on the J8 connector of the cg1000-RTP (Figure 3-5): Figure 3-5. J7 and J8 pinouts - rear of analyzer. iouT1 + Pin 3 iouT1-...
  • Page 28: Alarm Contact Connections

    A service alarm event triggers these alarm contacts to an open condition. Oxygen/Flow Alarm AlARM 3A Pin 12 Connections are configurable in AlARM 3B Pin 13 software to activate under a vari- AlARM 4A Pin 14 ety of user defined conditions, AlARM 4B Pin 15 3-8 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 29: Rs-232 Communications Connections

    RS-232 Communications Connections The CG1000-RTP communicates with other computers using RS-232 serial communications protocol. The connections used for RS-232 communica- tions are on connector J7 and labeled as follows (see Figure 3-1 for location of J7 on back of CG1000-RTP). Pin 1 Pin 2 Pin 3...
  • Page 30: Start-Up And User Interface

    The following topics are included: • System Start-up • Password Restrictions • Front Panel Display • Auto Ranging Feature • Keypad Keys • Selecting Menu Options • Exiting Menu Options • Automatic Menu Exit • On-Line Help 3-10 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 31: System Start-Up

    System Start-Up Plug the power cord into the power entry module on the back of the analyzer. Plug the other end of the power cord into the appropriate Ac power outlet. Turn on the analyzer by pressing on the side of the power switch marked 1.
  • Page 32: Keypad Keys

    To cancel out of a specific menu, press Cancel. To cancel out of all menus, continue to press Cancel until all menus disappear from the display and only process readings remain. Pressing Cancel does not delete any settings you have set while in  that menu. NOTE 3-12 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 33: Automatic Menu Exit Feature

    Automatic Menu Exit Feature If you don’t press a key, after 15 minutes the system exits the menu you were in and returns the system to normal operations (exits out of all menus). On-Line Help When you select a menu, you can press Help to get a description of the function of any of the menu options within that menu.
  • Page 34 This page intentionally left blank. 3-14 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 35: Chapter 4 User Interface

    If you don’t want to enable a system pass- word, skip this section. If you cannot remember the correct password, call AMETEK/Dycor  and a service password will be provided so you can access the Pass- word option and enter a new system password.
  • Page 36: Setup Key

    Flow Sometimes menu options can occupy Display Lines 3 and 4, and even  Display Line 2. Therefore, put the information that is most critical to view on Display Line 1 and 2. NOTE 4-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 37: Setting Up The Display

    Setting Up the Display 1. From the Setup menu, select Display. 2. Select Display Line #1, Display Line #2, or Display Line #3. 3. Select the type of information you want to place on the display line. Entering user text User text can be up to 20 characters long.
  • Page 38: Set Pressure

    Serial Output port until stopped by the user. This is used to test communication settings and wiring. This is used to view internal software variables Read Memory Location used for system operation and troubleshooting. 4-4 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 39: System Serial

    System Serial # View your system serial # and manufacturing #. Service personnel may ask for this information. 1. From the Setup menu, select System Serial # . 2. Select System Serial # or Manufacturing #. 3. The system displays the system serial # or manufacturing #. Primary Calibration A thermal calibration is not a regular calibration.
  • Page 40: Communication

    3. Select the carrier gas used to deliver gases to the analyzer. Choices are nitrogen, carbon dioxide, argon, or helium. If measuring close to 100% oxygen, we recommend that you select nitrogen as the carrier gas. 4-6 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 41 Calibrate flow sensor The flow sensor is calibrated at the factory. Only calibrate the flow sensor  if the analyzer loses its EEPROM memory. Always use a mixture of NOTE nitrogen and oxygen as your carrier gas when calibrating the flow sensor. You will need a mechanical 1.
  • Page 42: Analog Range Key

    Decide whether to enable output filtering. If enabled, output Filtering decide how much to smooth out abrupt changes in readings. Define the type of information the analog output should track. select Function cell Temperature Thermocouple Millivolts cell Millivolts 4-8 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 43: Analog Output Port

    Analog Output Port Before you can define any Analog Range Key functions, you must  choose whether it will apply to Analog Output Port 1 or 2 on the back of the analyzer. Analog Output Port 1 is labeled I OUT 1; Analog NOTE Output Port 2 is labeled I OUT 2.
  • Page 44: Set Track/Hold Option

    For calibration operations, select one of the following: Hold during cal Track during cal Select one of these options, your choice will be confirmed on the display: Out #1 will TRACK during Calibration Out #1 will HOLD during calibration 4-10 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 45: Set Current Mode Option

    For verify operations, select one of the following: Hold during verify Track during verify Select one of these options, your choice will be confirmed on the display: Out #1 will TRACK during Verify Out #1 will HOLD during Verify Set Current Mode Option Define either a 0-to-20 or 4-to-20 mA analog output range for the selected analog output port.
  • Page 46: Output Filtering Option

    The CG1000-RTP prompts you with the Set Current Range menu option where you can choose the range of readings that the 0 to 20 mA or 4 to 20 mA analog output represents (see the Set Current Range section in this chapter for details). 4-12 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 47: Alarm Key

    Alarm Key Alarm Key Menu Functions/Options Define alarm set point values. Wheth- er these alarms are set to activate Alarm set Points for high- or low-alarm conditions is defined using the Alarm Configure menu option. Define whether each oxygen alarm Alarm Configure (Alarms 3 and 4) should be a high or low alarm.
  • Page 48: Overview Of Alarm Functions

    Alarm 4 for a high/high O alarm, and the O value exceeds both the Alarm 3 and Alarm 4 set points, two up arrows will appear next to the O value on the display. 4-14 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 49 Defining alarm set points 1. From the Alarm menu, select Alarm Set Points. Once you select the Alarm Set Points menu option, the system alerts  you that Alarms 1 and 2 are reserved for the watchdog and service alarms, and are not software selectable. Press Enter to continue to the NOTE configurable alarms.
  • Page 50: Alarm Configure

    2. Select De-energize on Alarm or Energize on Alarm. If you select Energize on Alarm, the following appears: Relays are set to ENERGIZE on Alarm If you select De-Energize on Alarm, the following appears: Relays are set to DE-ENERGIZE on Alarm 4-16 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 51: Select Function

    Select Function Select the type of reading that Alarm 3 should track: • Oxygen/Flow • When the system is performing a calibration • When the system is performing a verification If tracking the start of a calibration or verification, the alarm will deacti- vate once the calibration or verification has been completed, including any recovery duration you have specified.
  • Page 52: Watchdog Relay

    20 most recent entries in this exception log. Pressing the Down Arrow Key shows later exception log entries. Pressing the Up Arrow Key shows more recent exception log entries. Press Cancel at any time to exit this menu option. 4-18 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 53: Chapter 5 Calibrate Key

    CALIBRATE KEY Overview Calibrate Key Function/Options initiate cal start a calibration initiate verify start a verify cal/verify data view previously collected calibration or verify data cal gas values Enter calibration gas values to match known values of calibration gas cylinders Configure Verify Enter verify gas value Enter acceptable verify range...
  • Page 54: Calibration

    The zero gas is the low calibration gas. For verify operations only one gas is used.  NOTE Aborting a Calibration You can cancel a calibration or verification at any time by pressing the Cancel key. 5-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 55: Initiate Cal

    Initiate Cal Perform a calibration. If you replaced a cell, or your cell has failed a regular calibration  because of a deterioration, you will need to perform a primary calibra- tion rather than a regular calibration. NOTE Enter the calibration gas value to match the zero calibration gas cylinder ...
  • Page 56 12. Set the 3-way valve on the back of the analyzer to GAS IN.  When returning to reading the sample gas, be sure to again adjust the flow rate to 150 sccm (± 5 sccm). NOTE 5-4 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 57: Initiate Verify

    Initiate Verify Perform a verification to check that the system is within specification. Before you can initiate a verify, you must do the following: Enter the verify gas value to match the verify gas cylinder you intend to  use to verify your system (Configure Verify). nter the verification gas value to match the verification gas you use to verify your system (see the Configure Verify section later in this chapter ...
  • Page 58 9. Set the 3-way valve on the back of the analyzer to GAS IN. When returning to reading the sample gas, be sure to again adjust the  flow rate to 150 sccm (± 5 sccm). NOTE 5-6 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 59: Cal/Verify Data

    Cal/Verify Data View the results of the latest calibration or verification. For calibration data, this includes the following: • Span calibration gas value vs. span calibration gas reading. • Zero calibration gas value vs. zero calibration gas reading. • Span calibration gas drift. • Zero calibration gas drift. For verify data, the display shows information on the verify gas. Span Value or Zero Value is the value the system reads from the cal gas. The term Span Drift or Zero Drift is the difference between the entered calibration gas value and the calibration gas values read by the system.
  • Page 60 If the date or time has not been set, random characters will appear  instead of the time and date when the calibration was completed. If this occurs, reset the analyzer’s time and date. If this doesn’t resolve NOTE the problem, a calibration is needed. 5-8 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 61: Cal Gas Value

    Cal Gas Value Define the zero calibration gas value to match the calibration gas cylinder you use to calibrate your system. • Span Gas - high calibration gas • Zero Gas - low calibration gas The terms Span Gas and Zero Gas used in this sec- The Span Gas value you enter must be higher than the Zero Gas tion refer to the calibration ...
  • Page 62: Configure Verify

    Alarm will trip. This mes- range is between 11% and 9%. sage will be cleared when a successful verify is run (the verify gas is within the range specified), or a calibration is run. 5-10 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 63: Set Clock

    Set Clock Set current time and date and define a recovery time. Set Time and Date Set the date or time for the analyzer’s internal clock. 1. From the Calibrate key menu, select Set Clock. 2. Select Set Time & Date from the Set Clock menu option. The analyzer uses military time so 16:00 equals 4:00 PM., The time is: YY:XX...
  • Page 64: Recovery Duration

    Enter new time: Mn:Sc output or triggering of alarms. 3. Enter the recovery time, then press ENTER. If you want to use the currently displayed recovery time, press ENTER without entering any text. 5-12 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 65: Chapter 6 Maintenance And Troubleshooting

    MAINTENANCE AND TROUBLESHOOTING The operations in this chapter should be performed only by qualified service personnel with a knowledge of electrical safety techniques. Never service the analyzer unless power has been disconnected. System and Error Messages System and error messages appear on the fourth line of the display at three-second intervals until the condition is corrected or has ended.
  • Page 66: System Messages

    The system is performing a Primary Calibration. The message con- Primary Calibration tinues to be displayed during the one-hour recovery period after the thermal calibration. System Calibrating System calibration is in progress. System verification is in progress. System Verifying Cal 6-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 67: Error Messages

    Error Messages Error messages on the control unit display indicate problems with the op- eration of the analyzer. Identify the type of problem indicated by the error message and go to “Diagnostic Checks” at the end of this chapter for help on how to pinpoint the problem.
  • Page 68 - Incorrect flow rates between the span gas and calibration gases - Calibration gas value incorrectly entered - Failed cell (thermal calibration should be run at this time) - Leak or plug in plumbing 6-4 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 69: Diagnostic Checks

    Diagnostic Checks Display the thermocouple millivolts and cell millivolts during trou-  bleshooting as a troubleshooting aid. You may need this information if you need to contact the factory for assistance. Be sure to always NOTE include your analyzer model and serial number when calling the fac- tory for technical support.
  • Page 70: Calibration Checks

    The best calibration gas to use for this check is O zero of your analyzer leading to lower- calibration gas. than-expected oxygen readings. 6-6 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 71: Ac Power Checks

    AC Power Checks Loss / Inadequate AC Voltage to the Sensor Measure the AC voltage to the backplane board at J4, Pins 1 and 2. Ensure that this voltage is sufficient. AC Power Checks Check the measurement technique used by the volt meter (for example, RMS, average, peak, etc.).
  • Page 72: General Troubleshooting

    Cell + and Cell - con- nections on the backplane board to see if the cell millivolts change, indicating a leak. Do not spray gas on the reference side of the cell. 6-8 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 73: Chapter 7 Parts Replacement

    PARTS REPLACEMENT The operations in this chapter should be performed only by quali- fied service personnel. There are no operator-serviceable components inside the CG1000-RTP, and an operator should never open the cover of the analyzer. This appendix shows you how to replace system parts and is divided into the following sections: •...
  • Page 74: Cell Replacement

    7. Pull the cell from the mount nut in the panel and remove from the analyzer. Discard the cell O-ring. 8. If you remove the ceramic inlet tube, check that it is not cracked. If cracked, order a new one. Figure 7-1. Cell location. 7-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 75 4. Install the ceramic inlet tube into the flat-faced union. Apply special grease (AMETEK P/N 06058JE) to the #6 O-ring and to the flat face of the flat-faced union. Tighten the union into the cell tee by hand, then snug the fitting with a 7/16”...
  • Page 76 9. Place the top cover onto the analyzer. 10. Tighten the screws that connect the top cover to the analyzer. 11. Reattach the power plug to the analyzer. 12. Turn on the analyzer. 13. Recalibrate the analyzer. 7-4 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 77: Furnace Replacement

    Furnace Replacement disconnect the power cord from the analyzer when servicing system  components. High voltages are present inside the analyzer when power is applied. The furnace is hot, even after power has been removed from the ana-  lyzer. use caution and wear appropriate gloves if removing the furnace insulation cover to access the furnace.
  • Page 78 Push the thermocouple stop against the Swagelok nut, then tighten with a wrench one-quarter turn. Place the new furnace into the furnace L-bracket; tighten the furnace cap- tive screw. Make sure that the furnace sits against the “L” shape bracket. 7-6 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 79 As you put the furnace insulation cover back on the furnace, place the thermocouple and furnace wires attached to the furnace through the hole in the furnace insulation cover. Reinstall the plate that fastens the furnace insulation cover to the fur- nace by reversing the actions taken in Step 5 of removing the furnace.
  • Page 80: Replacing The Transformer

    (in that order) on the transformer and tighten. 8. Reconnect J3, J4, and J7 connectors. 9. Put the top cover back on the analyzer. 10. Reapply power to the analyzer. 11. Recalibrate the analyzer. 7-8 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 81: Replacing The Backplane Board

    Replacing the Backplane Board Disconnect the power cord from the analyzer when servicing sys- tem components. High voltages are present inside the analyzer when power is applied. The furnace insulation cover and furnace are extremely hot, even after power has been removed from the analyzer. Use caution and wear ap- propriate gloves if removing the furnace insulation cover to access the furnace.
  • Page 82: Replacing Display/Keypad Modules

    If you are replacing the display module. When you receive the new module, you must also remove jumpers  AI2, AI3, and AI4 before reinstalling the display module. You can leave jumper AI5 in place. NOTE 7-10 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 83: Appendix A Serial Communications

    SERIAL COMMUNICATIONS Communication Protocol Communicating to the CG1000-RTP requires the following parameter set- ting: 8 Data bits 1 Stop bit No Parity In addition, the baud rate on the CG1000-RTP analyzer must match the baud rate of the host device. Baud rates are user-programmable. See the “Communication”...
  • Page 84: Command Message Format

    ?? (?? is equivalent to 2 ASCII 3F hex characters). This makes testing by the programmer easy but it is strongly recom- NOTE mended that the integrity check is enables in the final system. A-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 85: General Serial Communication Commands

    General Serial Communication Commands This section provides general serial communication commands. Program- ming commands and factory setup commands are provided later in this chapter. Normally you will use these general serial communication com- mands. Read Number (F) Read the value of a variable stored in the CG1000-RTP. The data section of this command is a number from 00-FF.
  • Page 86: Development Serial Communication Commands

    Due to the nature of EEPROM memory, each location in the memory has a limited number of writes allowed (approximately 10,000). The location character can be used to determine where the data is stored internally to the CG1000-RTP and act accordingly. A-4 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 87: Command List

    Command List This section provides the command letters you can use. Command Letter Description Echo Bad command Acknowledge Read number calibrate write number data Format Read date and time set date and Time Figure A-1. Serial communication commands list. Serial Communications...
  • Page 88: Responses

    Description Bad command letter Bad checksum input overrun in serial communication Parameter out of range Error in receipt of character cannot calibrate/verify now internal error Figure A-2. illegal access (read-only variable) Failure code list. A-6 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 89: Variable Table

    Variable Table COMMAND VARIABLE NAME FORMAT DATA FORMAT LETTER Flags_3 0 - sensor at op. temp 1 - over/under temp 2 - inject cal gas Active 3 - Reserved 4 - Reserved 5 - Reserved 6 - Reserved 7 - cal/verify not Permitted inf_mess_flag XX XX XX XX Byte 1 (bit #)
  • Page 90 14 - Flow display line #2 same as command letter 05 display line #3 same as command letter 05 oxygen FloAT cell Temp FloAT cell mv FloAT T/c mv FloAT Figure A-3. Main variable table (2). A-8 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 91 COMMAND VARIABLE NAME FORMAT DATA FORMAT LETTER current output #1 Function 0 - oxygen 2 - cell Temperature 4 - T/c mv 6 - cell mv current output #2 Function same as command letter 0E current output #1 20 mA sP FloAT current output #1 4 mA sP FloAT...
  • Page 92 % (see command 1E for ppm) Alarm 4 sP FloAT in % (se command 1F for ppm) Flow FloAT Hi_flow_set FloAT Flow High Alarm sP lo_flow_set FloAT Flow low Alarm sP Figure A-3. Main variable table (4). A-10 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 93: Appendix B Menu Option Charts

    MENU OPTION CHARTS Setup Key Menu Setup Key Menu Menu Options cell Millivolts Flow date/Time Display Thermocouple Millivolts cell Temperature user Text Turn off Password Enter password Set Pressure set Process Pressure System Tests various serial # System Serial # Manufacturing # Thermal Calibrate Perform a thermal calibration...
  • Page 94: Calibrate Key Menu

    High and low Flow Energize on alarm Relay Configure de-Energize on alarm Flow Select Function start of calibration (applies to Alarm 3 only) start of verify Exception Log Figure B-3. Alarm Key menu. B-2 | Dycor CG1000-RTP Oxygen Analyzer...
  • Page 95: Analog Range Key Menu

    Analog Range Key Menu Analog Range Key Menu Menu Options 20 mA Set Current Range 0 or 4 mA Hold output at last process reading during calibration Set Track/Hold Track calibration readings during calibration 4 to 20 mA Set Current Mode 0 to 20 mA Enter output filter value Output Filtering...
  • Page 96 This page intenionally left blank. B-4 | Dycor CG1000-RTP Oxygen Analyzer...

Table of Contents