Table of Contents

Advertisement

Responses

Figure A-2.
Failure code list.
A-6 | Dycor CG1000-RTP Oxygen Analyzer
The CG1000-RTP returns a variety of responses that can be broken into
two categories,
Success
Success
All successful responses start with
low the
and have a checksum after it. All responses end with a carriage
A
return ( ).
Success
End
A
Success
Data
A
(n characters)
Examples:
A 
Simple acknowledge
AFF0050
Acknowledge with data
Failure
All failure responses start with a
and a carriage return ( ).
Failure Code
Failure
(2 hex ASCII characters)
N
Failure Code
01
02
03
05
08
09
0A
0B
and
.
Failure
. If there is data returned, it will fol-
A
Checksum
(2 hex ASCII characters)
followed by a two hex digit failure code
N
End
Description
Bad command letter
Bad checksum
input overrun in serial communication
Parameter out of range
Error in receipt of character
cannot calibrate/verify now
internal error
illegal access (read-only variable)
End
Character