Figure 7-3.
Replacing the cell.
7-4 | Dycor CG1000-RTP Oxygen Analyzer
The exhaust fitting portion of the cell tee should be oriented at 9
o'clock as you look in from the front of the analyzer. Do not tighten
the cell tee to orient it properly - loosen the cell mount nut to reposi-
tion the exhaust tube.
6. Reconnect the exhaust tubing elbow to the cell tee. Hand tighten the
Swagelok nut, then use a wrench to tighten an additional one-quarter
turn. Always use a backup wrench when working on sensor plumb-
ing.
7. Reinstall the inlet and exhaust plumbing by reversing earlier steps.
8. Place the cell clip on the cell using a pair of long-nosed pliers (Figure
7-3).
9. Place the top cover onto the analyzer.
10. Tighten the screws that connect the top cover to the analyzer.
11. Reattach the power plug to the analyzer.
12. Turn on the analyzer.
13. Recalibrate the analyzer.