Cal/Verify Data

View the results of the latest calibration or verification.
For calibration data, this includes the following:
• Span calibration gas value vs. span calibration gas reading.
• Zero calibration gas value vs. zero calibration gas reading.
• Span calibration gas drift.
• Zero calibration gas drift.
For verify data, the display shows information on the verify gas.
Span Value or Zero Value is the value the system reads from the cal gas.
The term Span Drift or Zero Drift is the difference between the entered
calibration gas value and the calibration gas values read by the system.
Both the latest calibration and verification data will be stored in memory.
Setting Up Cal/Verify Data Option
1. From the Calibrate Key menu, select Cal/Verify Data.
2. Select Calibration Data or Verify Data.
NOTE
Calibration data
For calibration data, the display first shows the span gas calibration data.
3. Press Enter to view the zero gas values.
4. Press Enter.
The time when the calibration was completed displays:
If there are minor differences between the values and what the difference
is, it is because the system truncates numbers when displaying them.
Span Value: xx
Span Gas:
Span Drift:
Zero Value: xx
Zero Gas:
Zero Drift:
End time hr: mn
End date mm/dd/yr
xx
xx
xx
xx
Calibration
|
5-7