Alarm Contact Connections - Ametek Dycor CG1000-RTP User Manual

Table of Contents

Advertisement

Watchdog Timer
Default: Deactivated (Closed)
A watchdog event triggers these alarm contacts to an open condition. If you have a watchdog
event, and the system returns to normal operations, the watchdog alarm will reset. The system
will retain in memory the last time the watchdog condition occurred.
Service Alarm
Default: Deactivated (Closed)
A service alarm event triggers these alarm contacts to an open condition.
Oxygen/Flow Alarm
Connections are configurable in
software to activate under a vari-
ety of user defined conditions,
3-8 | Dycor CG1000-RTP Oxygen Analyzer
Powering standard current outputs from an external voltage
The J8 connector on the back of the CG1000-RTP (Figure 3-5) contains a
set of pins to power the standard current outputs using an external volt-
age source:
VEXT
VEXTCOM
Jumper
Define which power source to use.
JP2
JP2 is in
Analyzer uses the internal supply to power the current output
(factory
circuits. if you try to use an external voltage at this time, the
default)
system will use whichever voltage is higher
CG1000-RTP voltage is normally about 28 Volts.
JP2 is
External voltage source must be used to power the cur-
removed
rent output circuits. The external voltage source must be no
greater than 30 volts dc.
To determine the minimum external voltage supply required to power the
current outputs, use the following formula:
VEXTmin = .02 amps (RLoad + Rwires) + 4 volts
VEXTmin = minimum external voltage required to power current output
circuit
RLoad =
load resistance of the current output device
Rwires =
resistance of the wires

Alarm Contact Connections

Connector J8 (Figure 3-5) also contains alarm pinouts.
AlARM 1A
AlARM 1B
AlARM 2A
AlARM 2B
AlARM 3A
AlARM 3B
AlARM 4A
AlARM 4B
Pin 9
Pin 10
Pin 5
Pin 6
Pin 7
Pin 8
Pin 12
Pin 13
Pin 14
Pin 15