Perform a calibration.
NOTE
1. From the Calibrate Key menu, select Initiate Cal.
2. Select Continue Cal to continue, or Abort Cal to cancel the calibration
and enter the values or set up calibration plumbing.
You are prompted to run the O
3. Set the 3-way valve on the back of the analyzer to the CAL position.
4. Set the 3-way valve that delivers the calibration gases to the SPAN
position.
5. Press ENTER once you have applied the span gas. The span gas dis-
play appears.
6. Adjust the flow rate to 150 sccm. When the O
of the first display is stable, press ENTER.
NOTE
If you replaced a cell, or your cell has failed a regular calibration
because of a deterioration, you will need to perform a primary calibra-
tion rather than a regular calibration.
Enter the calibration gas value to match the zero calibration gas cylinder
you are using to calibrate your system.
Have you entered
the Cal/Vrfy values?
Abort Calibration
Inject Span\Verify Gas
ENTER to continue
CANCEL to Abort
O2 xx.x %
Press ENTER to Span
The reading on the far right of the first display line shows the setpoint
value you entered using the Cal Gas Values menu option. This set-
point value should match the cal gas cylinder for the O
Continue Cal
span gas.
2
Set xx.x%
Flow x sccm
reading on the far left
2
span gas.
2
|
Calibration
5-3