Ametek Dycor CG1000-RTP User Manual page 90

Table of Contents

Advertisement

COMMAND
VARIABLE NAME
LETTER
03
04
05
06
07
08
0B
0c
0d
Figure A-3.
Main variable table (2).
A-8 | Dycor CG1000-RTP Oxygen Analyzer
iout_flags
Probe type
display line #1
display line #2
display line #3
oxygen
cell Temp
cell mv
T/c mv
FORMAT
HEX
inT
inT
inT
inT
FloAT
FloAT
FloAT
FloAT
DATA FORMAT
Analog output Flags
XX XX
1 2
Byte 1 (bit#)
- Analog output #2
7 - Future use
6 - 0 = (4-20) ma
1 = (0-20) ma
5 - 0 = Holds during verify
1 = Tracks during verify
4 - 0 = Holds during cal
1 = Tracks during cal
- Analog output #1
3 - Future use
2 - 0 = (4-20) ma
1 = (0-20) ma
1 - 0 = Holds during verify
1 = Tracks during verify
0 - 0 = Holds during cal
1 = Tracks during cal
Byte 2 (bit#)
- Analog output #2
7 - Future use
6 - 0 = (4-20) ma
1 = (0-20) ma
5 - 0 = Holds during verify
1 = Tracks during verify
4 - 0 = Holds during cal
1 = Tracks during cal
- Analog output #1
3 - Future use
2 - 0 = (4-20) ma
1 = (0-20) ma
1 - 0 = Holds during verify
1 = Tracks during verify
0 - 0 = Holds during cal
1 = Tracks during cal
same as command letter 05
0 - Blank line
2 - oxygen
4 - cell Temperature
6 - cell Millivolts
10 - current Time and date
12 - user Text
14 - Flow
same as command letter 05
same as command letter 05