Calibration

5-2 | Dycor CG1000-RTP Oxygen Analyzer

Calibration vs. Verification Operations

during a verify operation, a calibration gas reading is taken, and the
difference between what the system records for the gas value and
Verify
its known value (from the gas cylinder) is recorded. Internal calibra-
tion parameters are not changed. verify operations are often used to
check that the instrument is running within required tolerances.
during a calibration operation, software algorithms adjust the dif-
Calibration
ference between the current readings and the known calibration
gas values until they are equal.

Span vs. Zero Gas

This chapter also refers to the terms span and zero calibration gases. The
span gas is the high calibration gas. The zero gas is the low calibration gas.
For verify operations only one gas is used.
NOTE

Aborting a Calibration

You can cancel a calibration or verification at any time by pressing the
Cancel key.