4-16 | Dycor CG1000-RTP Oxygen Analyzer
Alarm Configure
Define whether alarms should trigger based on high oxygen, low oxygen,
or combined high/low flow condition.
Alarm 3 is set high or low oxygen or combined high/low flow. Alarm 4
is set to trigger high or low oxygen.
NOTE
Configuring the alarms
1. From the Alarm menu, select Alarm Configure.
2. Define Alarm 3 as a high oxygen, low oxygen, or combined high/low
flow and Alarm 4 as high oxygen or low oxygen.
3. Press Cancel to exit this menu option.
To create two high oxygen alarms, define both Alarm 3 and Alarm 4 as
high oxygen alarms.
Relay Configure
All Series CG1000-RTP relays are Form A—normally open contacts. You
can use the software to set whether the alarms should close on an alarm
condition (energize) or open on an alarm condition (de-energize).
Configuring the relay
1. From the Alarm menu, select Relay Configure.
2. Select De-energize on Alarm or Energize on Alarm.
If you select Energize on Alarm, the following appears:
If you select De-Energize on Alarm, the following appears:
Relays are set to
ENERGIZE on Alarm
Relays are set to
DE-ENERGIZE on Alarm