Ametek Dycor CG1000-RTP User Manual page 60

Table of Contents

Advertisement

If during a verify operation,
the system finds that the verify
gas is not within the acceptable
range, a Verify Failure message
will appear on the bottom line
of the display, and the Service
Alarm will trip. This message
will be cleared when a successful
verify is run (the verify gas is
within the range specified), or a
calibration is run.
5-8 | Dycor CG1000-RTP Oxygen Analyzer
If the date or time has not been set, random characters will appear
instead of the time and date when the calibration was completed. If
this occurs, reset the analyzer's time and date. If this doesn't resolve
NOTE
the problem, a calibration is needed.
Verify Data
For verify data, the display shows the verify gas data:
The term verify value on the display is the value the system read during
the verify operation, and the term verify gas is the set point value entered
using the Configure Verify option from the Calibrate Key menu.
• Press ENTER.
The time when the verification was completed appears on the
display:
If the date or time has not been set, random characters will appear
instead of the time and date when the calibration was completed. If
this occurs, reset the analyzer's time and date. If this doesn't resolve
NOTE
the problem, a calibration is needed.
Verify Value:
xx
Verify Gas:
xx
Verify Diff:
xx
End time hr:mn
End date mm/dd/yr