2. Select whether you want to enter the Alarm 3 set point using a percent
or PPM value. This option won't appear if Alarm 3 is not set to activate
based on an O
tion or verification.
3. You will be prompted to define the Alarm 3 set point value:
4. Enter the Alarm 3 set point value, then press Enter (just press Enter if
you don't want to change the current value).
5. Select whether you want to enter the Alarm 4 set point using a percent
or PPM range.
You are prompted to enter the Alarm 4 set point value:
6. Enter the Alarm 4 set point value.
7. Press Cancel or Enter to exit this menu option.

Alarm Configure

Define whether alarms should trigger based on a high-alarm or a low-alarm
condition.
1.
For Alarm 3, set whether to track high oxygen, low oxygen, or the
combined high/low flow.
2.
For Alarm 4, set whether to track high oxygen or low oxygen.
NOTE
Configuring the alarms
1. From the Alarm menu, select Alarm Configure.
2. Define Alarm 3 and Alarm 4 as high or low alarms.
3. Press Cancel to exit this menu option.
NOTE
reading, but is set to activate on the start of a calibra-
2
Alarm #3 = xx
New Value?
Alarm #4 = xx O2
New Value?
Whether Alarm 3 is set high or low only applies if it is set to track
oxygen. If Alarm 3 is set to trigger at the start of a calibration or veri-
fication, this option does not apply.
To create two high oxygen alarms, define both Alarm 3 and Alarm 4
aw high oxygen alarms.
Controller / User Interface
|
4-21