Commands And Error Messages Applicable To The Rs; Interface; Interface Write Commands (Data Issued To The Thermostat) - Lauda ECO Silver E 4 S Operation Manual

Table of Contents

Advertisement

F.7
Commands and error messages applicable to the RS 232/485 interface module and to the
Ethernet interface
F.7.1 Interface write commands (data issued to the thermostat)
OUT_PV_05_XXX.XX
OUT_SP_00_XXX.XX
OUT_SP_01_XXX
OUT_SP_02_XXX
OUT_SP_04_XXX
OUT_SP_05_XXX
OUT_PAR_00_XXX.X
OUT_PAR_01_XXX
OUT_PAR_02_XXX
OUT_PAR_03_XX.X
OUT_PAR_04_XX.XX
OUT_PAR_05_XXXX
OUT_PAR_06_XXXX
OUT_PAR_07_XXXX.X
OUT_PAR_09_XXX.X
OUT_PAR_10_XX.X
OUT_PAR_14_XXX.X
OUT_PAR_15_XXX
OUT_MODE_00_X
OUT_MODE_01_X
OUT_MODE_03_X
OUT_MODE_04_X
START
STOP
RMP_SELECT_X
RMP_START
RMP_PAUSE
RMP_CONT
RMP_STOP
RMP_RESET
RMP_OUT_00_XXX.XX_XXXXX_XXX.XX_X
RMP_OUT_02_XXX
114 / 124
Command
Specify external temperature via interface
Set-value transfer with max. 3 places before the decimal point and max. 2 places after it.
Pump power level 1 to 6
Cooling operating mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
TiH outflow temperature limit, upper value
TiL outflow temperature limit, lower value
Setting of the control parameter Xp.
Setting of the control parameter Tn (5 – 180 s; 181 = Off).
Setting of the control parameter Tv.
Setting of the control parameter Td.
Setting of the control parameter KpE.
Setting of the control parameter TnE (0 – 9000 s; 9001 = Off).
Setting of the control parameter TvE (5 = OFF).
Setting of the control parameter TdE
Setting of the max. correcting quantity limit.
Setting of the control parameter XpF.
Setting of the setpoint offset.
Setting of the control parameter PropE.
Keypad: 0 = released / 1 = locked (corresponds to: "KEY").
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analog / 3 = ext. serial.
Keypad Command remote control: 0 = released / 1 = locked.
Setpoint offset source: 0 = normal / 1 = ext. Pt / 2 = ext. analog / 3 = ext. serial.
Switches the device on (from Standby)
Switches the device in Standby (pump, heating, chiller off).
Selection of program (1 — 5) to which further commands are to refer. When the device is
switched on Program 5 is selected.
Start the programmer.
Stop the programmer.
Start the programmer again after a hold.
Terminate the program.
Delete program (all segments)
Sets programmer segment (temperature, time, tolerance, and pump level). A segment is
appended and assigned appropriate values.
Number of program loops: 0 = endless / 1 — 250.
ECO Silver