Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B, E4407B and E4408B
E7402A, E7403A, E7404A, and E7405A, Option 1DN
Tracking Generator Level Flatness:
E4402B, E4403B, E4404B, E4405B, E4407B and E4408B
E7402A, E7403A, E7404A, and E7405A, Option 1DN

Test Limits

Flatness ≤ 10 MHz
Flatness > 10 MHz

Test Description

A calibrated power sensor is connected to the tracking generator output to measure the
power level at 50 MHz. The power meter is set for REL mode so that future power level
readings are in dB relative to the power level at 50 MHz. The tracking generator is then
stepped to several frequencies throughout its range. The output power difference relative
to the power level at 50 MHz is measured at each frequency and recorded.
Item
Meters
Power Meter
RF Power Sensor
Frequency Range: 100 kHz to 3.0 GHz
Adapters
Type-N (f) to BNC
(m), 75-Ohm
Type-N (m to BNC (f)
(2)
Cables
(2) BNC, 122-cm
(48-in)
BNC, 23-cm (9-in)
190
Maximum
Maximum
−3.5 dBm
+3.5 dBm
−2.5 dBm
−2.5 dBm
Critical Specifications
(for this test)
Recommended
Agilent Model
438A or E4418A,
E4419A
8482A
1250-1534
1250-1476
10503A
10502A
Chapter 4