PSA Functional Tests
Frequency Readout Accuracy

Procedure

1. Configure the equipment as shown in
alignment has been performed within the past 24 hours.
2. Perform the following steps to set up the equipment:
a. On the synthesized sweeper, press
POWER LEVEL
, 1505,
CW
MHz
b. On the spectrum analyzer, press
3. Set up the spectrum analyzer by pressing:
,
System
Reference
, 1505,
Frequency
,
Det/Demod
Detector
, 2990,
Span
MHz
Ensure Ref Level is set to 0 dBm. In addition, ensure Resolution BW and
NOTE
Video BW are both set to Auto.
4. Press
on the analyzer to measure the frequency readout accuracy. If the
Peak Search
instrument is functioning correctly, the marker reading in the active function block will
be between the values listed in
Frequency Readout column in
5. On the spectrum analyzer, change the span and center frequency as listed in
6. Change the synthesized sweeper frequency to match the center frequency of the
analyzer.
7. Repeat steps 4 through 6 until the Marker Frequency Readout column of
complete.
Table 3-4
Frequency Readout Accuracy Results
Span
Center
(MHz)
Frequency
(MHz)
2990
1505
127.2
1505
54.1
1505
7.95
1505
0.106
1505
1.98
517.59
1.98
832.50
128
Figure
PRESET
, –10,
dBm
,
(On)
RF
.
Preset
, 10,
,
(On)
MHz
Ext
MHz
,
Sample
Table
3-4. Record the marker value in the Marker
Table
3-4.
Minimum
Marker Frequency Readout
1.495 GHz
1.5045 GHz
1.50480 GHz
1.504968 GHz
1.5049996 GHz
517.5829 MHz
832.4928 MHz
3-4.. Confirm the analyzer's built-in auto
, then set the controls as follows:
Table
3-4.
Table 3-4
is
Maximum
1.515 GHz
1.5055 GHz
1.50520 GHz
1.505032 GHz
1.5050004 GHz
517.5971 MHz
832.5071 MHz
Chapter 3