Figure 3-1. Amplitude Accuracy Test Setup
Procedure
1. Zero and calibrate the power meter.
2. Configure equipment as shown in
spectrum analyzer input.
To minimize stress on the test equipment connections, support the power
CAUTION
sensor.
3. If the auto alignment for the analyzer has not been performed within the past 24 hours,
press
,
System
Alignments
4. Press
(Factory Preset) on the analyzer.
Preset
5. Set up the spectrum analyzer by pressing:
, 50,
Frequency
MHz
, 2,
Span
MHz
,
Input/Output
RF Coupling
,
Amplitude
Attenuation
,
Sweep
Auto Swp Time
6. Set up the synthesized sweeper by pressing:
, 50,
CW
MHz
, -4,
Power Level
dBm
(On)
RF
Chapter 3
Figure
3-1., with the power splitter connected to the
,
to perform the auto alignment routine.
Align All Now
(DC)
, 10,
dB
,
Accy
PSA Functional Tests
Amplitude Accuracy at 50 MHz
121