Procedure (501 Mhz To 1.0 Ghz) E4401B And E4411B E7401A - Agilent Technologies PSA Series Manual

Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

4. Press
. Record the marker frequency next to your analyzer model in the
Peak Search
Measured Frequency column as entry (a) or (b) in
500 MHz.
5. Press the following keys on the analyzer:
,
(Cont)
Sweep
Sweep
,
BW/Avg
Average
,
(Auto)
BW/Avg
Res BW
(Auto)
Video BW
, 50,
SPAN
kHz
FREQUENCY
6. Press
, and set the center frequency of the analyzer to the frequency recorded
Center Freq
in the Measured Frequency column as entry (a) or (b) of
10 MHz to 500 MHz.
7. Press the following keys on the analyzer:
,
, 1,
BW/Avg
Res BW
, 30,
Video BW
Hz
Single
Wait for the sweep to finish.
8. Press the following keys on the analyzer:
,
Display
Display Line
Adjust the display so that it is centered on the average trace noise, ignoring any
residual responses.
9. Record the display line amplitude setting as TR Entry 1 (TR Entry 19 or 20 for a 75 Ω
Input) in
Table 4-1 on page
Maximum for the appropriate input impedance.

Procedure (501 MHz to 1.0 GHz) E4401B and E4411B E7401A

1. Press the following keys on the analyzer:
,
(Cont)
Sweep
Sweep
,
FREQUENCY
Start Freq
, 1.0,
Stop Freq
GHz
,
, 1,
BW/Avg
Res BW
, 10,
Video BW
kHz
2. Press the following keys on the analyzer:
Single
,
BW/Avg
Average
Single
Wait until AVG 3 is displayed to the left of the graticule (the analyzer will take 3
sweeps, then stop).
Chapter 4
(Off)
kHz
(On)
157. The Average Noise Level should be less than the
, 501,
MHz
MHz
(On), 3,
Enter
ESA Functional Tests
Displayed Average Noise Level
Table 4-1 on page 157
for 10 MHz to
Table 4-1 on page 157
for
143