Figure 4-8 Equipment Setup

Procedure

1. Zero and calibrate the power meter and power sensor as described in the power meter
operation manual.
2. Connect the equipment as shown in
Use only 75 Ω cables, connectors, or adapters on analyzers with 75 Ω
CAUTION
connectors, or damage to the connectors will occur.
3. Set the synthesized sweeper controls as follows:
,
FREQUENCY
Center Freq
, –8,
POWER LEVEL
dBm
4. Press
,
System
Power On/Preset
the preset routine to finish. Set the analyzer by pressing the following keys.
, 50,
FREQUENCY
, 50,
CF Step
MHz
, 20,
SPAN
kHz
,
AMPLITUDE
More
, –10,
AMPLITUDE
,
AMPLITUDE
Attenuation
, 2,
Scale/Div
dB
,
, 10,
BW/Avg
Res BW
, 3,
Video BW
kHz
Peak Search
,
FREQUENCY
Signal Track
5. Adjust the synthesized sweeper power level for a marker amplitude reading of –14 dBm
+/– 0.10 dB.
Chapter 4
Figure 4-7
, 50,
MHz
,
(Factory),
Preset Type
MHz
,
,
Y Axis Units
dBm
dBm
, 10,
dB
kHz
(On)
Frequency Response (Flatness)
or
Figure
4-8.
on the analyzer and wait for
Preset
ESA Functional Tests
169