Amplitude Linearity; Test Limits; Test Description - Agilent Technologies PSA Series Manual

Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Amplitude Linearity

Test Limits

The linearity error will be ≤ ±1.0 dB with ≤ −10 dBm at the mixer.

Test Description

This test checks the amplitude linearity of the instrument by maintaining a constant
reference level and measuring signals of different amplitudes over most of the display
range. This test sets the input attenuator to 10 dB and the Reference Level to 0 dBm. The
external attenuator is set to 0 dB, and the amplitude of the source is adjusted to set the
displayed signal at the reference level.
The instrument's internal marker is used to measure the reference amplitude. The Marker
Delta function is activated and the RF input is reduced using the external precision step
attenuator. Signal input levels from 0 dBm to −50 dBm are measured.
Item
Adapter
Type-N(m), to BNC(f)
Adapter
3.5 mm(f) to 3.5mm(f)
Adapter
BNC(m) to SMA(f)
APC 3.5 mm (1
APC 3.5 mm, 1 meter
meter)
Attenuator, 10 dB
Range: 0-50 dB
Step
Frequency: 50 MHz
Accuracy: ±0.25 dB
Cables (2 required)
BNC, 120 cm (48 in.)
Output Level Accuracy: 0 to –15 dBm: ±1.0 dB
Synthesized
Sweeper
Chapter 3
Critical Specifications
(for this test)
PSA Functional Tests

Amplitude Linearity

Recommended
Agilent Model
1250-1476
5061-5311
1250-1700
8120-4921
355D
10503A
83630B
133