Frequency Response (Flatness); Test Limits; Test Description - Agilent Technologies PSA Series Manual

Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Frequency Response (Flatness)

Test Limits

ESA EMC Model
E4411B & E4401B
E7401A
E4402B & E4403B
E7402A
E4404B
E7403A
E4405B
E7404A
E4407B & E4408B
E7405A

Test Description

The output of the synthesized sweeper is fed through a power splitter to a power sensor
and the analyzer as shown in
level is adjusted at 50 MHz to place the displayed signal at the analyzer center horizontal
graticule line. Measurements are made at various points depending on the model being
tested. The signal source amplitude is measured with a power meter to eliminate errors
due to source flatness. The power meter is zeroed and calibrated before starting the
measurement.
Item
Signal Sources
Synthesized
Sweeper
Adapters
Chapter 4
Frequency
9 kHz to 1.5 GHz
9 kHz to 3 GHz
9 kHz to 3 GHz
3.01 GHz to 6.7 GHz
9 kHz to 3 GHz
3.01 GHz to 6.7 GHz
6.71 GHz to 13.2 GHz
9 kHz to 3 GHz
3.01 GHz to 6.7 GHz
6.71 GHz to 13.2 GHz
13.21 GHz to 25 GHz
25 GHz to 26.5 GHz
Figure 4-7
or
Figure
Critical Specifications
(for this test)

Frequency Response (Flatness)

Minimum
Maximum
(dB)
(dB)
–1.5
1.5
–1.5
1.5
–1.5
1.5
–3.0
3.0
–1.5
1.5
–3.0
3.0
–3.5
3.5
–1.5
1.5
–3.0
3.0
–3.5
3.5
–4.0
4.0
–4.5
4.5
4-8. The synthesized sweeper's power
Recommended
Agilent Model
8340A/B or 836xx Series
ESA Functional Tests
167