ESA Functional Tests
Displayed Average Noise Level
3. Press
. Record the marker frequency next to your analyzer model in the
Peak Search
Measured Frequency column as entry (q) (m) in
13.2 GHz.
4. Press the following keys on the analyzer:
,
(Cont)
Sweep
Sweep
BW/Avg, Average
,
(Auto)
BW/Avg
Res BW
(Auto)
Video BW
, 50,
SPAN
kHz
FREQUENCY
5. Press
. Set the center frequency of the analyzer to the frequency recorded in
Center Freq
the Measured Frequency column as entry (q) (m) in
to 13.2 GHz.
6. Press the following keys on the analyzer:
,
, 1,
BW/Avg
Res BW
, 30,
Video BW
Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
,
Display
Display Line
Adjust the display line so that it is centered on the average trace noise, ignoring any
residual responses.
8. Record the display line amplitude setting as TR Entry 15 13 in
The Average Noise Level should be less than the Maximum.

Procedure (12.01 GHz to 22 GHz) E4407B and E4408B E7405A

1. Press
,
System
Power On/Preset
preset routine to finish. Set the analyzer by pressing the following keys:
,
FREQUENCY
Start Freq
, 22,
Stop Freq
GHz
,
AMPLITUDE
More
, –70,
AMPLITUDE
(Man), 0,
Attenuation
,
, 1,
BW/Avg
Res BW
, 10,
Video BW
kHz
154
(Off)
kHz
(On)
,
(Factory),
Preset Type
, 12.01,
GHz
,
,
Y Axis Units
dBm
dBm
dB
MHz
Table 4-1 on page 157
for 12.01 GHz to
Table 4-1 on page 157
Table 4-1 on page
on the analyzer. Wait for the
Preset
for 12.01 GHz
157.
Chapter 4