Linear Mode Procedure - Agilent Technologies PSA Series Manual

Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Table 4-5
Reference Level Accuracy Worksheet (Log Mode) (Continued)
Analyzer
Reference Level
Sweeper Amplitude
dBm
dBmV
–40
11.76
Reference + (–10 dB)
–50
1.76
Reference + (–20 dB)
–60
– 8.24
Reference + (–30 dB)
–70
–18.24
Reference + (–40 dB)

Linear Mode Procedure

1. Set the power level on the synthesized sweeper to –30 dBm by pressing
–30,
.
dBm
2.
Set the analyzer by pressing the following keys:
,
(Cont)
Sweep
Sweep
,
AMPLITUDE
More
, –30,
AMPLITUDE
,
AMPLITUDE
More
, +21.76,
AMPLITUDE
(Lin)
Scale Type
3. Adjust the amplitude on the synthesized sweeper until the marker amplitude on the
analyzer reads –30 dBm +/– 0.10 dB. Enter the synthesized sweeper power level as the
Synthesized Sweeper Amplitude reference in
NOTE
Under these analyzer conditions, the sweep time is 1.7 seconds. Therefore,
the marker amplitude updates are fairly slow when adjusting the synthesizer
output power.
4. Now that the reference has been established in step 4, adjust the synthesized sweeper
power level and the analyzer reference level according to
synthesized sweeper output power and the analyzer's reference level will be changed in
10 dB steps.)
5. On the analyzer, press
,
.
Marker
Delta
6. For each new synthesized sweeper power level and analyzer reference level change,
press the following keys on the analyzer:
Single
Peak Search
Chapter 4
Synthesized
Minimum
(dBm)
–11.40
–21.40
–31.40
–42.0
(50 Ω Input)
,
,
Y Axis Units
dBm
(50 Ω Input)
dBm
(75 Ω Input)
,
,
Y Axis Units
dBmV
(75 Ω Input)
dBmV
, wait for a sweep to finish, and then press
Single
Reference Level Accuracy
Analyzer
Marker ∆
(dB)
Amplitude
(dB)
Table 4-6 on page
176.
Table 4-6 on page 176
ESA Functional Tests
Maximum
(dB)
–8.60
–18.60
–28.60
–38.0
,
Power Level
(The
,
Peak Search
175