Procedure (12.01 Ghz To 13.2 Ghz) E4405B E7404A - Agilent Technologies PSA Series Manual

Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

4. Press the following keys on the analyzer:
,
(Cont)
Sweep
Sweep
BW/Avg, Average
,
(Auto)
BW/Avg
Res BW
(Auto)
Video BW
, 50,
SPAN
kHz
FREQUENCY
5. Press
. Set the center frequency of the analyzer to the frequency recorded in
Center Freq
the Measured Frequency column as entry (o) or (p) (l) in
GHz to 12.0 GHz.
6. Press the following keys on the analyzer:
,
, 1,
BW/Avg
Res BW
, 30,
Video BW
Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
,
Display
Display Line
Adjust the display line so that it is centered on the average trace noise, ignoring any
residual responses.
8. Record the display line amplitude setting as TR Entry 13 or 14 12 in
page
157. The average noise level should be less than the Maximum.

Procedure (12.01 GHz to 13.2 GHz) E4405B E7404A

1. Press
,
Preset System
Wait for the preset routine to finish. Set the analyzer by pressing the following keys:
,
FREQUENCY
Start Freq
, 13.2,
Stop Freq
GHz
,
AMPLITUDE
More
, –70,
AMPLITUDE
(Man), 0,
Attenuation
,
, 1,
BW/Avg
Res BW
, 10,
Video BW
kHz
2. Press the following keys on the analyzer:
Single
,
BW/Avg
Average
Single
Wait until AVG 3 is displayed to the left of the graticule (the analyzer will take 3 sweeps
and then stop).
Chapter 4
(Off)
kHz
(On)
,
Power On/Preset
Preset Type
, 12.01,
GHz
,
,
Y Axis Units
dBm
dBm
dB
MHz
(On), 3,
Enter
ESA Functional Tests
Displayed Average Noise Level
Table 4-1 on page 157
Table 4-1 on
(Factory),
, on the analyzer.
Preset
for 6.01
153