6. Press the following keys on the analyzer:
,
, 1,
BW/Avg
Res BW
, 30,
Video BW
Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
,
Display
Display Line
Adjust the display line so that it is centered on the average trace noise, ignoring any
residual responses.
8. Record the display line amplitude setting as TR Entry 11 10 in
The Average Noise Level should be less than the Maximum.

Procedure (6.01 GHz to 6.7 GHz) E4404B E7403A

1. Press
,
System
Power On/Preset
preset routine to finish. Set the analyzer by pressing the following keys:
,
FREQUENCY
Start Freq
, 6.7,
Stop Freq
GHz
,
AMPLITUDE
More
, –70,
AMPLITUDE
(Man), 0,
Attenuation
,
, 1,
BW/Avg
Res BW
, 10,
Video BW
kHz
2. Press the following keys on the analyzer:
Single
,
BW/Avg
Average
Single
Wait until AVG 3 is displayed to the left of the graticule (the analyzer will take 3 sweeps
and then stop).
3. Press
. Record the marker frequency next to your analyzer model in the
Peak Search
Measured Frequency column as entry (n) (k) in
6.7 GHz.
4. Press the following keys on the analyzer:
,
(Cont)
Sweep
Sweep
BW/Avg, Average
,
(Auto)
BW/Avg
Res BW
(Auto)
Video BW
, 50,
SPAN
kHz
FREQUENCY
Chapter 4
kHz
(On)
,
(Factory),
Preset Type
, 6.01,
GHz
,
,
Y Axis Units
dBm
dBm
dB
MHz
(On), 3,
Enter
(Off)
Displayed Average Noise Level
Table 4-1 on page
on the analyzer. Wait for the
Preset
Table 4-1 on page 157
for 6.01 GHz to
ESA Functional Tests
157.
151