Tracking Generator Level Flatness: E4402B, E4403B, E4404B, E4405B, E4407B and E4408B
Figure 4-19 Equipment Setup
Figure 4-20 Equipment Setup
Procedure
1. Calibrate the tracking generator by pressing
Connect the RF OUT to the RF INPUT when prompted.
2. Press
,
System
Power On/Preset
for the preset routine to finish. Set the analyzer by pressing the following keys:
,
FREQUENCY
Center Freq
, 100,
CF Step
MHz
,
SPAN
Zero Span
,
AMPLITUDE
More
Source, Amplitude
,
System
Alignments
3. Zero and calibrate the power meter and power sensor. Make sure the power meter is
reading out in dBm. Enter the power sensor 50 MHz cal factor into the power meter.
4. Connect the power sensor to the RF Out on the analyzer as shown in
Figure
4-20.
5. Press REL on the power meter. The power meter readout amplitudes are now relative to
the power level at 50 MHz.
Chapter 4
,
(Factory),
Preset Type
, 50,
MHz
,
,
Y Axis Units
dBm
(On), –20,
dBm
,
,
Auto Align
Off
E7402A, E7403A, E7404A, and E7405A, Option 1DN
,
,
System
Alignments
Align Now
on the analyzer, then wait
Preset
ESA Functional Tests
,
.
TG
Figure 4-19
or
191