Procedure (1.01 Ghz To 1.5 Ghz) E4401B And E4411B E7401A - Agilent Technologies PSA Series Manual

Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

4. Press the following keys on the analyzer:
,
(Cont)
Sweep
Sweep
,
BW/Avg
Average
,
(Auto)
BW/Avg
Res BW
Video
(Auto)
BW
, 50,
SPAN
kHz
FREQUENCY
5. Press
. Set the center frequency of the analyzer to the frequency recorded in
Center Freq
the Measured Frequency column as entry (c) (b) in
to 1.0 GHz.
6. Press the following keys on the analyzer:
,
, 1,
BW/Avg
Res BW
, 30,
Video BW
Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
,
Display
Display Line
Adjust the display line so that it is centered on the average trace noise, ignoring any
residual responses.
8. Record the display line amplitude setting as TR Entry 2 (TR Entry 21 for a 75 Ω Input)
in
Table 4-1 on page
the appropriate input impedance.

Procedure (1.01 GHz to 1.5 GHz) E4401B and E4411B E7401A

1. Press the following keys on the analyzer:
,
(Cont)
Sweep
Sweep
,
FREQUENCY
Start Freq
,
Stop Freq
1.5 GHz
,
, 1,
BW/Avg
Res BW
, 10,
Video BW
kHz
2. Press the following keys on the analyzer:
Single
,
BW/Avg
Average
Single
Wait until AVG 3 is displayed to the left of the graticule (the analyzer will take 3
sweeps, then stop).
3. Press
, and record the marker frequency next to your analyzer model in the
Peak Search
Measured Frequency column as entry (d) or (e) (c) in
to 1.5 GHz.
Chapter 4
(Off)
kHz
(On)
157. The Average Noise Level should be less than the Maximum for
, 1.0
GHz
MHz
(On), 3,
Enter
ESA Functional Tests
Displayed Average Noise Level
Table 4-1 on page 157
for 501 MHz
Table 4-1 on page 157
for 1.01 GHz
145