Second Harmonic Spurious Responses
Test Limits
Model Number
Maximum
E4401B
–55 dBc
E7401A
E4402B
–55 dBc
E7402A
E4403B
–50 dBc
E4404B
–55 dBc
E7403A
E4405B
–55 dBc
E7404A
E4407B
–55 dBc
E7405A
E4408B
–50 dBc
E4411B
–55 dBc
Test Description
To test second harmonic distortion, a 50 MHz low pass filter is used to filter the source
output, ensuring that harmonics read by the analyzer are internally generated and not
coming from the source. The source power and input attenuation on the analyzer are
adjusted so –20 dBm is the power level at the first mixer.
Item
Signal Sources
Synthesized Sweeper
Miscellaneous
50 MHz Low pass
Rejection at 80 MHz: >60 dB
filter
Adapters
(2) Type-N (m) to
BNC (f)
BNC (f) to BNC (f)
Chapter 4
Critical Specifications
(for this test)
ESA Functional Tests
Second Harmonic Spurious Responses
Recommended
Agilent Model
8340A/B or 836XX
Series
0955-0306
1250-1476
1250-0080