ESA Functional Tests
Tracking Generator Level Flatness: Models E4401B and E4411B E7401A, Options 1DN and
1DQ
Figure 4-17 Equipment Setup
Figure 4-18 Equipment Setup

Procedure

1. Calibrate the tracking generator by pressing
Connect the RF Out to the Input when prompted.
2. Press
,
System
Power On/Preset
for the preset routine to finish. Set the analyzer by pressing the following keys:
,
FREQUENCY
Center Freq
, 500,
CF Step
MHz
,
SPAN
Zero Span
,
AMPLITUDE
More
Source, Amplitude
,
Source
Amplitude
3. Zero and calibrate the power meter and RF power sensor. Make sure the power meter is
reading out in dBm. Enter the power sensor 50 MHz cal factor into the power meter.
75 Ω RF Out only: Zero and calibrate the 75 Ω power sensor.
NOTE
4. Connect the power sensor to the RF Out on the analyzer as shown in
Figure
4-18.
188
,
(Factory),
Preset Type
, 50,
MHz
,
,
Y Axis Units
dBm
(50 Ω RF Output only)
(On), 0,
dBm
(75 Ω RF Output only)
(On), 42.76,
dBmV
,
,
System
Alignments
Align Now
on the analyzer, then wait
Preset
,
.
TG
Figure 4-17
or
Chapter 4