Procedure - Agilent Technologies PSA Series Manual

Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Figure 4-16 Equipment Setup

Procedure

1. Connect the equipment as shown in
2. Set the synthesized sweeper controls as follows:
, 40,
Frequency
MHz
, –10,
POWER LEVEL
, –4.3,
POWER LEVEL
75 Ω Input only. Connect the minimum loss adapter between the low pass
NOTE
filter and 75 Ω Input.
3. Press
,
System
Power On/Preset
preset routine to finish.
Set the analyzer by pressing the following keys:
,
AMPLITUDE
More
,
FREQUENCY
Center Freq
, 1,
SPAN
MHz
, –10,
AMPLITUDE
, 44,
AMPLITUDE
dBmV
(Man), 10,
Attenuation
,
, 30,
BW/Avg
Res BW
4. Adjust the synthesized sweeper power level to place the peak of the signal at the
reference level.
5. Set the analyzer by pressing the following keys:
, 50,
SPAN
kHz
,
, 1,
BW/Avg
Res BW
, 100,
Video BW
Hz
Chapter 4
Figure 4-15
(50 Ω Input only)
dBm
(75 Ω Input only)
dBm
,
(Factory),
Preset Type
,
,
Y Axis Units
dBm
, 40,
MHz
(50 Ω Input only)
dBm
(75 Ω Input only)
dB
kHz
kHz
Second Harmonic Spurious Responses
or
Figure
4-16.
on the analyzer. Wait for the
Preset
ESA Functional Tests
185