7.11.3 Internal and external control parameter sets

If a thermostat is used for a number of applications, which always leads to a change of the control parameters, these
control parameters (up to 9 sets) can be saved in the thermostat and activated again as required.
Also saving is useful for finding the best control parameters; in this way external management of the control
parameters can be avoided.
There are 9 sets (each for internal and external sets of control parameters) saved at the factory.
In this menu the control parameters cannot be edited, they are only displayed.
With
Activate
With
Upload actual
With
Default
set by the customer are lost).
Command
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Correction limitation
Pump
Menu
Status
intern
extern
Pump
Menu
84 / 137
the currently valid control parameters are used.
the actual ones are read in and saved (for later reuse).
the set of control parameters saved at the works is loaded again (in this case the control parameters
Set 1
Set 2
Set 3
Set 4
Set 5
Set 6
Set 7
Set 8
Set 9
End
T
set
Activate
Upload actual
Default
End
T
set
Proline Kryomats
Control parameter sets
 Open the device parameter menu via the soft
key
 With the cursor keys continue to:
Control
Parameters
 The adjacent window appears.
Set 1 to Set 9.
 Select the desired set with
with
T
fix
 Select the desired set to be changed with
and confirm with
 In the setting window (see left) the selected set
is listed under
display.
 Under
is activated,
is read in and
the set, which was saved at the factory, is
restored.
T
fix
Menu .
Control
Control para. sets
and confirm
.
.
internal
or
external
Status
the previously selected set:
.
in the
03/2022