Internal And External Control Parameter Sets - Lauda Proline RP 845 C Operating Instructions Manual

Table of Contents

Advertisement

Proline Edition X cooling thermostats

7.13.3 Internal and external control parameter sets

If a thermostat is used for a number of applications, which always leads to a change of the control
parameters, these control parameters (up to 9 sets) can be saved in the thermostat and activated again
as required.
Also saving is useful for finding the best control parameters; in this way external management of the
control parameters can be avoided.
There are 9 sets (each for internal and external sets of control parameters) saved at the factory.
In this menu the control parameters cannot be edited, they are only displayed.
With
Activate
With
Upload actual
With
Default
control parameters set by the customer are lost).
Command
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Pump
Menu
Status
intern
extern
Pump
Menu
YACE0100 / October 4, 2018
the currently valid control parameters are used.
the actual ones are read in and saved (for later reuse).
the set of control parameters saved at the works is loaded again (in this case the
Set 1
Set 2
Set 3
Set 4
Set 5
Set 6
Set 7
Set 8
Set 9
End
T
set
Activate
Upload actual
Default
End
T
set
Control parameter sets
Open the device parameter menu via the
soft key
With the cursor keys continue to:
Control para. sets
The adjacent window appears.
Set 1 to Set 9.
 Select the desired set with
confirm with
 Select the desired set to be changed with
T
fix
In the setting window (see left) the
selected set is listed under
external
Under
set:
is activated,
is read in and
the set, who was saved at the factory, is
restored.
T
fix