4.3

Filling with heat transfer liquid and draining

V3
WARNING!
Overflow of heat transfer liquid
Electric shock
Do not overfill the device. Monitor the liquid level in the
tank.
WARNING!
Spraying of heat transfer liquid
Electric shock
Avoid spraying heat transfer liquid.
CAUTION!
Risk of heat transfer liquid escaping due to the use of unsuitable
hoses
Scalding, cold burns
The temperature and media resistance of the hoses must be
suitable for the application.
NOTICE!
Use of unsuitable heat transfer liquids
Device damage
In accordance with the descriptions of the LAUDA heat
transfer liquids in this operating manual, only the heat
transfer liquids of water and water/glycol are approved.
The heat transfer liquid must be chosen based on the tem-
perature range of its application.
Note the following instructions:
The device is designed exclusively for non-flammable heat transfer liq-
uids.
LAUDA is not liable for damages resulting from the use of unsuitable
heat transfer liquids.
Only connect pressure-sealed consumers to the device.
Once the consumer is connected, the liquid level in the tank will sink as
the consumer is filled. Top up the heat transfer liquid!
When consumers are higher up, the external volume can run empty
when there is a standing pump and intrusion of air into the thermostatic
circuit, even in a closed circuit. This will cause the tank in the device to
overflow!