Filling and emptying

LAUDA accepts no liability for damage caused by the use of unsuitable heat transfer liquids (approved heat transfer liq-
uids ( 6.4)).
Caution!
Caution!
Notice
Filling
Withdraw the drain tap.
-
Optimum operation is ensured with a filling level of 20 - 40 mm below the bath bridge (max. filling level: 20 mm).
-
Operation is possible down to a filling level of 60 mm below the bath bridge; a low level alarm occurs from a filling
-
level of approx. 90 mm below the bath bridge. ( 8.1)
When using oils as heat transfer liquids note that they expand on heating (approx. 10 % per 100 °C).
-
Take into account the displacement volume of any objects to be introduced into the bath.
-
With a connected external consumer the complete expansion takes place in the bath.
-
09/2021
Contact with heat transfer liquid when filling / draining
Harmful when inhaled,
damage to eyes and skin
Pay attention to the safety data sheet for the heat transfer
liquid.
Use CE gloves, protective clothing and eye protection during
physical contact with heat transfer liquid.
Avoid splashing the heat transfer liquid.
Make sure that the drain tap is closed before filling.
Use of unsuitable heat transfer liquid
Explosion, burns, scalds, fire
When selecting the heat transfer liquid, observe the admissi-
ble temperature range.
Only use LAUDA heat transfer liquids.
Overfilling containers, spilling heat transfer liquid
Environmental hazard from leaking heat transfer liquid
Note the thermal volume expansion of the heat transfer liq-
uid.
Where necessary, consider the displacement volume of the
body being introduced.
Take the volume of external consumers into account.