System does not start completely if collimator is in
manual mode
Details
Cause
Brief Solution
System does not start completely if collimator is in
manual mode. An error message is displayed indicating
a problem with the collimator during start up.
The key on collimator was not switched back to auto-
matic mode. The system checks communication with
all components during start up. If collimator is in man-
ual mode no communication with the system is availa-
ble.
Set the key at the backside of the collimator to auto-
matic.
Restart the system on X-ray generator console. A restart
of NX is not needed.