Unloading of the bucky in the Radiographic Wall
Stand
To unload the bucky with a cassette or detector:
1. Open the bucky drawer completely by pulling the handle.
2. Rotate the carrier back to portrait position.
3. Push firmly with both hands the cassette or detector towards the lower
clamp to open the clamping mechanism.
4. Remove the cassette or detector from the clamping. The openings in the
carrier allow your fingers to grip the detector or cassette.
5. Load the bucky with another cassette or detector.
• Alternatively, close the bucky drawer using the front handle and
pushing the button to release the brake.
CAUTION:
Make sure your fingers are not between slider and cassette.
The clamping mechanism may hurt your fingers, therefore
take special care.