Error numbers

Display
Error
Action
tube kV max
1
B
control A max
2
B
tube mA max
3
B
tube +-kV diff
4
B
tube +-mA diff
5
B
ROM test
6
A
RAM test
7
A
unknown
8
B
no Tube kV
9
B
tube kV too high
10
B
inverter overload
11
B
send timeout
12
B
E²Prom checksum
13
A
DR 400 | X-Ray Generator Mini Console | 197
Explanation
Tube voltage too high (>166 kV/132 kV)
Load current too high (> 250 A)
Tube current too high (> 900 A)
Tube voltage difference between +URist
and- URist>15kV
Tube current difference between +IRist
and -IRist >100mA
ROM test checksum error
RAM test error
unknown error
Tube voltage < 10kV after 1ms or < 50%
after 30ms
Tube voltage > rated voltage + 25%
Converter overload ( > 150000WS)
Serial interface transmit timeout
E2PROM checksum error
3231E EN 20210708 1530