96 | DR 400 | Basic Workflow
For example if your adult abdomen settings are: 70--85 kVp, 200--400 mA,
15--80 mAs, consider starting at 65--75 kVp, 100--160 mA, 2.5--10 mAs for a
pediatric patient. Whenever possible use high kVp techniques and large SID
(Source Image Distance).
Summary:
• Image only when there is a clear medical benefit.
• Image only the indicated area.
• Use the lowest amount of radiation for adequate imaging based on size of
the child (reducing tube output -- kVp and mAs).
• Try to use always short exposure times, large SID values and immobilizing
devices.
• Avoid multiple scans and use alternative diagnostic studies (such as
ultrasound or MRI) when possible.