The standard position of the X-ray tube arm is indicated by markers. When the
tube arm is in standard position it is centered in transversal direction on the
bucky.
Related Links
Movement ranges
Manual Collimator Technical Data
Automatic Collimator Technical Data
Positioning the Bucky
Centering and collimating
Topics:
Stop positions
Collision indicator
DR 400 | Radiographic Table and X-Ray Tube Stand | 133
Rotation of the collimator around the axis of the X-ray beam
Note: To avoid shock and damage, move the column with
normal speed and slow down when reaching the mechanical
end stops.
CAUTION:
If a grinding noise is heard during vertical movement of the X-
ray tube arm or radiographic wall stand, the steel cables inside
the tube stand or wall stand could be broken. Do not operate the
unit any further and try to avoid hard vibration or knocks of any
kind. Please contact service.
CAUTION:
Rotation may be limited by cables. Avoid strain on the cables
during rotation.
on page 234
on page 139
on page 173
on page 244
on page 245
3231E EN 20210708 1530