200 | DR 400 | X-Ray Generator Mini Console
Error
Action
42
B
43
A
44
B
45
B
46
C
47
A
48
B/C
49
B
50
B
51
A
52
A
53
B
61
B
62
B
3231E EN 20210708 1530
Display
grid is on
RTC checksum, Batt.
low
starter system
no main current
exposure stopped by
user
controller - E²prom ver-
ify
Wrong tube position
Tube mA too high
Device not ready
(CAN)
No BUS-Signal from
AEC
FLXIS not ready
Anode heat content
>100% !
Receiver overflow
Transmitter overflow
Explanation
Grid active in idle state
RTC (real time clock) error, RTC battery
discharged
Normal speed starter fault current
Load current < 4A after 0.5 ms
Exposure aborted by operator
Cpu-E2Prom alignment
Tube position sensor
Tube current outside tolerance
Device or device interface not ready
No stop signal from automatic exposure
control
No communication to TV chain
Max Tube heat storage capacity, cool
down the tube
Serial interface receive buffer overflow
Serial interface transmit buffer overflow