AGFA DR 400 User Manual page 198

Hide thumbs Also See for DR 400:
Table of Contents

Advertisement

198 | DR 400 | X-Ray Generator Mini Console
Error
Action
14
B
15
B
16
A
17
B
18
A
19
A
20
B
21
B
22
B
23
B
24
C
25
B
26
B
27
A
3231E EN 20210708 1530
Display
watchdog
receive timeout
E²Prom wait timeout
fillament system
DAP system
filament parameter
+-15V low
+5V low
key is on
XRAY key is on
mAs max
exposure too short
generator not ready
service intervall
Explanation
Watchdog error
Serial interface receipt timeout
E2PROM access timeout
Heater fault
Area dose measuring system self test error
Deviating heating parameters in E2PROM
+-15V outside tolerance
+5V outside tolerance
Key on control panel has been pressed
during switch-on
Exposure or fluoroscopy key has been
pressed during switch-on
Current time product in mAs has reached
its limit
Not used
Tube switchover error
Service interval for maintenance

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5520/1005520/200

Table of Contents