72 | DR 400 | Basic Workflow
1. Close the collimator blades fully.
Make sure that the collimator blades are fully closed and that no patient is
present in the room. To avoid radiation on a panel, remove the panel, turn
the tube away from the panel, or cover the panel with a lead apron.
2. Make sure that no one will be exposed.
3. On the software console, go to the screen with modality controls.
Figure 25: Navigation button for modality controls
4. Click the button to start the automated workflow for warming-up of the X-
ray tube.
Figure 26: Start the automated workflow for warming-up of the X-ray
tube
A table is displayed with a list of exposures.
1
2
1.
2.
Figure 27: List of exposures for warming-up of the X-ray tube
5. Perform the exposures and wait for the timer icon to finish between
exposures.
The exposure parameters are set automatically.
3231E EN 20210708 1530
Table with list of exposures
Button to cancel the warming-up procedure