Table 19: Automatic filter
S
Density
These buttons are used to adjust the AEC cut-off dose (and patient entrance
dose accordingly).
Density can be increased and decreased in a range of -4 to +4. Each step is a
change of one exposure step. An exposure step is a change of approximately
-20% or +25% in dose. When disabled, the density range number appears in
black.
Table 20: Dose variation compared to reference dose
Density
-4
-3
-2
-1
0
+1
+2
+3
+4
Patient Size
The size of the patient is classified in five categories: Extra Small, Small,
Medium, Large and Extra Large.
Touch the UP or DOWN arrows to select the desired patient size.
DR 400 | Software Console and Tube Head Display | 121
low dose
middle dose
high dose
Dose
0.41
0.51
0.64
0.80
1 (reference dose)
1.25
1.56
1.95
2.44
3231E EN 20210708 1530